Sensor Calibration - by SmartPrecision™ II Software
OPS
Series Encoders, Side and Top Mount Configurations
7.
Calibration
complete.
6.
Move the
sensor to the
Right Limit and
click the Set
button.
5.
Move the sensor
to the Left Limit
and click the Set
button.
8.
“Calibration Complete” message will disappear
after a few seconds, ending the calibration mode.
SmartPrecision II will automatically revert to the
mode in use prior to calibration.
Note: It is important that if you choose to calibrate only the limits, the sensor must be
placed in the “no limit” area of the scale (neither the left or right limit) when you click the
“Start Cal” button.